Troubleshooting

4.1
Troubleshooting
Troubleshooting
The following is list of troubleshooting tips and messages that you may encounter. They
include explanations, and in some cases, a recommended action. If a problem seems
unsolvable, contact your local Siemens office or regional Ultrasonic Flow Representative for
expert help at http://www.automation.siemens.com/partner
(http://www.automation.siemens.com/partner).
Table 4- 1
Troubleshooting Tips
Error or Message
Memory Full!
Memory Corrupted!
Chan Not Setup
Clr Active Memory?
Clr Saved Data?
<EOT>
No Sites - Press <ENTER>
Security
RTC Error
FUH1010 IP65 NEMA 7 Interface Detector Quick Start
Operating Instructions, 06/2014, A5E02428148-AA
Probable Cause
Response to an attempt to save site
data, when data memory is full.
Memory read error occurred while
accessing the active site data.
Response to an attempt to invoke an
operation that requires a channel to be
enabled.
Use this function to restore operation if
a severe event (e.g., a violent power
surge) disrupts system operation.
[Clr Saved Data?] only appears after
pressing <Down Arrow> in response to
[Clr Active Memory?].
Response to a request to output
Datalogger data to the printer or the
Graphics screen when no Datalogger
data exists or at the end of a
transmitted file.
Response while trying to recall/delete a
site setup when no sites are stored.
Response upon changing previously
entered data when security switch is in
[Disable] position or security code has
been entered.
Component level problem.
Solution
Delete an obsolete site or clear
Datalogger memory to make room for
the new data.
Refer to F4 reset procedure in the
Operation Instructions manual.
Enable the channel [Channel Setup -
Channel Enable - Yes]. Note that a
channel cannot be enabled until an
"Install" operation is completed.
Refer to F4 reset procedure in the
Operation Instructions manual.
Answering Yes to [Clr Saved Data?]
will erase ALL saved data. To invoke in
RS-232 serial mode, type @@@ and
then press <Enter> key.
Set up the Datalogger.
Create a site.
Change switch position to [Enable].
Enter previously set security code.
Meter requires service. Request
RMA.