Recommended separation distances for portable and mobile RF communication equipment and the EUT
The EUT is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the EUT can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the
EUT as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Explanation of utilized symbols
The following symbols are used on the device or on the packaging:
The CE marking indicates that the product complies with
Directive 93/42/EEC concerning medical devices.
Catalogue number
REF
Serial number
SN
Product bearing this symbol may not be disposed of together
with general household waste, but instead requires separate
disposal.
Temperature limitation for storage and transport.
Maximum humidity for storage and transport.
Maximum pressure for storage and transport.
Type BF applied part
Please read and follow these instructions for use and keep
them for future reference. (Background: blue, foreground: white)
Manufacturer
Production date
Interference may occur in the vicinity of equipment marked
with the following symbol.
Handle with care!
Keep dry!
"Grüner Punkt" (country-specific)
10
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 3,5/3 * SQRT (P)
0.1
0.4
1.2
3.7
11.7
80 MHz to 800 MHz
d = 3,5/3 * SQRT (P)
0.1
0.4
1.2
3.7
11.7
800 MHz to 2,5 GHz
d = 7/3 * SQRT (P)
0.2
0.7
2.3
7.4
23.3