If damaged, do not use this product!
Do not use in the presence of combustible or explosive materials!
Never lay this product on the patient or in his/her direct vicinity.
Operate the BiSect only with the Erbe connecting cable Art. No.
20196-106.
Protect this product from any form of mechanical damage! Do
not throw! Do not use force! If there is a cable, and unless stated
otherwise, do not kink it or wrap it around the accompanying
product!
Only transport, store, machine-clean or sterilize the product to-
gether with the protector (Art. No. 20195-194).
Erbe Elektromedizin expressly warns against modifying the
product. Any modification exempts Erbe Elektromedizin from
any and all liability.
4
Unit settings
4.1
Erbe ICC
• No special settings are needed.
• WARNING! Never activate a CUT mode.
• Deactivate AUTO START.
• Set the power limit according to the application.
4.2
Erbe VIO
• WARNING! Never activate a CUT mode.
• Deactivate AUTO START.
• Erbe recommends the BIPOLAR SOFT coagulation mode.
• Set the power limit according to the application.
24