OBJ_DOKU-175853-006.fm Seite 23 Donnerstag, 14. Juni 2018 11:06 11
IMPORTANT!
Please read all information carefully.
These instructions for use do not replace the user manual of the
electrosurgical unit used! Read the user manual of the electro-
surgical unit and ask Erbe or your distributor in case of doubt!
1
Normal Use
The BiSect (bipolar scissors) are used for:
• mechanical cutting with bipolar coagulation
• bipolar coagulation
• Preparation
of biological tissue in open surgery.
Erbe Elektromedizin recommends using this product on the fol-
lowing equipment:
• Erbe ICC
• Erbe VIO
2
Maximum electrical capacity
The maximum electrical capacity of this product is:
• 600 V
p
3
Safety instructions
This product may be used only by trained medical staff, who have
been shown how to use it according to the instructions for use.
Before every application, inspect the product and its entire insu-
lation (including the ceramic parts) for visible and mechanical
damage, and carry out a function test.
WARNING!
23