Code:
Meaning:
o
The main control board
has a memory problem.
Self help action:
Contact your dealer or Handicare.
Code:
Meaning:
r
No charge current.
Self help action:
Make sure that power supply is switched
on from mains. If it is contact your dealer
or Handicare.
Code:
Meaning:
y
The main control board
has a software fault.
Self help action:
Contact your dealer or Handicare.
Code:
Meaning:
.8.
N/A
Lift is in "sleep mode" to
conserve battery power.
Self help action:
Press toggle or handset button to wake lift.
Code:
Meaning:
P
P
Powered swivel
did not respond.
Self help action:
Contact your dealer or Handicare.
Code:
Meaning:
U
U
Faulty power supply.
Self help action:
Ensure the mains power supply is switched
on. If the fault continues, please contact
your dealer or Handicare.
Code:
Meaning:
N/A
Over current.
Self help action:
Contact your dealer or Handicare.