CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
3.3
FCB100, FCH100, FCB400 and FCH400 sensors in
remote mount design
NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
"write protection active" (see chapter 'Activating the
hardware write protection' on page 3 ).
Scope of delivery
1
2
Fig. 12
Item
Designation
1
Holding rod
2
Locking cylinder
Seal
Seal wire
Preparation
Fig. 13
1.
Unscrew the fixing screws
housing and replace with the holding rods
G12359
Quantity
1
1
1
1x
G12360
on the cover of the
3
2
.
Lay seal wire
1
Fig. 14
1.
Insert locking cylinder
housing cover.
2.
Insert the seal wire
holding rod
3.
Tightly twist the seal wire
cylinder
3
to secure the locking cylinder
NOTICE
You should still be able to slightly move the locking
cylinder after it is secured with the seal wire.
4.
Both ends of the seal wire must be very close together
and the twist loops should be smaller than 1 cm.
5.
Place the seal approx. 5 cm from the locking cylinder
onto the ends of the seal wire and press together using
sealing pliers. Cut off any excess seal wire.
1
2
3
2
3
on the holding rod
2
through the openings on the
1
such that both ends are the same length.
2
right behind the locking
EN – 9
3
G12352
1
on the
3
.
3

Werbung