Chapter 2
General
The instrument is equipped with four fuses:
-
Fuse F1 (1)
-
Fuse F2 (2)
-
Fuse F3 (3)
-
Fuse F4 (4)
The disassembly description below refers to fuse F1
(1).
Fuses F2 through F4 are disassembled the same way.
Disassembly - F1
Fuse cap (5) is equipped with a bayonet base.
Caution!
1. To open, first press cap inwards (using a screw
driver), then rotate 90 ° counterclockwise up to
limit stop and release.
Cap will be ejected together with fuse.
2. Remove fuse.
Disassembly - fuse holder
Unplug instrument from mains!
1. Disassembly - housing, upper right - see chapter
2.1.2
2. Disassembly housing, lower, right - see chapter
2.1.2.
3. Disconnect cable connection (6).
4. Loosen lock nut (7) from the inside of the housing.
5. Remove fuse holder (8) from instrument rear
panel.
Service Manual / Leica ST5020
Allgemeines
Im Gerät befinden sich vier Sicherungen:
-
Sicherung F1 (1)
-
Sicherung F2 (2)
-
Sicherung F3 (3)
-
Sicherung F4 (4)
Nachstehend wird die Demontage für die Sicherung
F1 (1) beschrieben.
Die Demontage für die Sicherung F2 bis F4 ist analog
dazu durchzuführen.
Demontage F1
Sicherungskappe (5) ist mit Bajonettverschluß verse-
hen.
Achtung !
1. Zum Öffnen Kappe mit Schraubendreher zuerst
eindrücken, danach ca. 90° entgegen dem
Uhrzeigersinn zum Anschlag drehen und loslas-
sen. Kappe wird zusammen mit Sicherung ausge-
worfen.
2. Sicherung entnehmen.
Demontage Sicherungshalter
1. Demontage Gehäuse-Oberteil-rechts,
siehe Kapitel 2.1.2.
2. Demontage Sockelblende rechts hinten,
siehe Kapitel 2.1.2.
3. Kabelverbindung (6) trennen.
4. Kontermutter (7) an Gehäuseinnenseite lösen.
5. Sicherungshalter (8) aus Gehäuserückwand
entnehmen.