Axio Observer materials
4.10.2
Changing the fuses in the Axio
Observer 7 materials
The Axio Observer 7 materials is supplied with
voltage from the external power unit VP232-2. The
fuse compartment of the VP232-2 power supply
unit is located at the rear of the unit and contains
two type T 4.0 A/H / 250 V fuses.
Pull out the power plug before
changing fuses.
•
Pull out the power plug.
•
Pull out the fuse holder (Fig. 13/2) frontward.
•
Remove the fuses from the fuse holder and
insert new fuses.
•
Reinsert
the
fuse
compartment (Fig. 13/1) until it clicks into place.
•
Plug in the power plug.
4.10.3
Changing the fuses in the ballast
unit for the HBO 100
The fuse holder for the F1 and F2 fuses is located
on the rear of the ballast unit. The fuse holder is
integrated into the power socket and contains two
T 2.0 A/H 250 V 5 x 20 mm fuses.
If the device fuses fail the cause must first of all be
ascertained
and
technical problems
remedied.
Pull out the power plug before
changing fuses.
•
Pull out the power plug.
•
Remove the fuse holder (Fig. 14/1) from the
fuse compartment (Fig. 14/2) by pulling it
towards you.
•
Replace the defective fuse.
•
Reinsert
the
fuse
compartment until it clicks into place.
•
Plug in the power plug.
09/2015
INSTALLATION NOTES AND FIRST-TIME SET-UP
holder
into
the
fuse
properly
holder
into
the
fuse
431004-8044-010
Fig. 13
Changing the fuses in the Axio
Observer 7 materials
Fig. 14
Changing fuses in the ballast unit
ZEISS
21