Herunterladen Inhalt Inhalt Diese Seite drucken

Read Fi Rst And Then; Scope Of Supply; Overview Of The Parts - Camille Bauer SINEAX V604 Betriebsanleitung

Universal-messumformer
Vorschau ausblenden Andere Handbücher für SINEAX V604:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Operating Instructions
Universal Transmitter SINEAX V 604
Contents
1. Read fi rst and then ... ................................................20
2. Scope of supply .........................................................20
3. Overview of the parts .................................................20
4. Brief description .........................................................21
5. Technical data ............................................................21
6. Exchanging frontplates ...............................................23
7. Withdrawing and inserting the device ........................23
8. Mounting ....................................................................23
9. Electrical connections ................................................24
10. Programming the transmitter .....................................26
11. Commissioning ...........................................................27
12. Maintenance ...............................................................27
13. Releasing the transmitter ...........................................27
14. Dimensional drawings ................................................27
15. Declaration of conformity ...........................................28

1. Read fi rst and then ...

The proper and safe operation of the device
assumes that the Operating Instructions
are read and the safety warnings given in
the various Sections
8. Mounting
9. Electrical connections
10. Programming the transmitter
11. Commissioning
are observed.
The device should only be handled by appropriately
trained personnel who are familiar with it and authorised
to work in electrical installations.
The instrument must only be opened for the setting of
the DIP switch, as described in section "10. Program-
ming the transmitter".
The guarantee is no longer valid if the instrument is
further tampered with.

2. Scope of supply

Transmitter (1)
Order Code: Signifi cance of the 2nd. and 3rd. digits
604-1 x x
1 Standard, measuring input not I.S.,
power supply 24... 60 V DC/AC
2 Standard, measuring input not I.S.,
power supply 85...230 V DC/AC
3 [EEx ia] IIC, measuring input I.S.,
power supply 24... 60 V DC/AC
4 [EEx ia] IIC, measuring input I.S.,
power supply 85...110 V DC / 85...230 V AC
2 Standard climatic rating;
instrument with cold junction compensation
4 Extra climatic rating;
instrument with cold junction compensation
20
ON
(
(
(1)
(6)
Fig. 1
2 withdrawing handles (2) (for withdrawing the device from
its housing)
2 Frontplates (3) (for notes)
2 Type labels (4) (for recording the operating data after
programming)
1 Ex approval (5) (only for Ex version devices)
1 Operating Instructions (6)

3. Overview of the parts

Figure 2 shows those parts of the transmitter of conse-
quence for mounting, electrical connections, programming
connections and other operations described in the Operating
Instructions.
SINEAX
V604
SINEAX
V604
ON
(
(
ON
(
(
(3)
(2)
No.
No.
(4)
(5)

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

987810-05

Inhaltsverzeichnis