Herunterladen Inhalt Inhalt Diese Seite drucken

Data Format; Output Data (Od) - Magtrol DSP7000 Betriebsanleitung

Inhaltsverzeichnis

Werbung

Magtrol DSP7000-Leistungsbremsen-Controller
7.2

DATA FORMAT

7.2.1
o
utput
OD (output data) Speed-torque data is a fixed-length string in ASCII format with a floating point
decimal. Use the following string format:
Where...
S = Speed in rpm. (No leading zeroes are used.)
d = Decimal digit 0 through 9.
T = Torque in units selected during setup. (The torque value always contains a decimal
L = Counterclockwise dynamometer shaft rotation (left).
R = Clockwise dynamometer shaft rotation (right).
. = Decimal point. (The decimal point location depends on the specific dynamometer
d
(od)
AtA
point.)
and torque range in use.)
Note:
When an "A" is in the R/L position (e.g. SdddddTdddd.A(cr)
(lf)), it is an indication that the unit is in an alarm condition.
Note:
The (cr) and (lf) characters will not display.
Kapitel 7 – Rechnergesteuerter Betrieb
SddddddTdddd.R(cr)(lf)
Or
SddddddTdddd.L(cr)(lf)
67

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Magtrol DSP7000