6.8. Active Energy (main display) + Time (secondary display)
Measurement
Warning
To avoid damages to the Meter or harms to you, do you
measure higher than AC voltage 750V rms and AC current
1000A rms.
To test for Active Energy (main display) + Time (secondary
display), connect the Meter as follows:
1. Dial Rotary to ENERGY range.
2. Press the lever to open the transformer jaw, and clamp them
to the corresponding phase of tested conductor. If user
needs to measure any phase of the 3 phase, then clamp
them to that phase's conductor.
3. The connecting method of 3 phase 4 wires and 3 phases 3
wires is possible
4. Press SEL to choose one of the three phases
* The double display shows the value of tested object's
active engergy kWh value and the measuring time of the
corresponding phase.
* The measuring reading gets increasing along with the time
increases. Press HOLD to read a particular time kWh
value. Then the reading and time are locked, but still
continuous accumulate measuring time.
* After read the data, press HOLD again to continous
measurement. kWh value continous accumulate and the
measuring time jumps to the present measuring time.