Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data; Measurement Tolerances - VOLTCRAFT Energy Logger 4000 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Energy Logger 4000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Error
No operation possible.
No data memor any more
available (MEM < 2%).
Repairs other than those just described should only be performed by an authorised spe-
cialist.
If you have queries concerning use of the measuring device, our technical support ser-
vice is available at the following telephone number:
Voltcraft®, 92242 Hirschau, Lindenweg 15,
Phone 0180 / 586,582,723 8

Technical data

Operating voltage
Max. power/current
Performance measurement display
Display energy use
Display
Tariff range
Accuracy
Buffer battery
Ambient conditions
Weight
Dimensions (LxWxH)
Overvoltage category
Pollution degree

Measurement tolerances

Statement of accuracy in ± (% of reading + display error in counts (= number of smallest points)). The accu-
racy is valid at a temperature of +23°C ± 5°C, and at a relative humidity of less than 75 %, non-condensing.
In no event exceed the max. permitted input values.
Possible cause
You are mode
„Data transmission".
The internal data memory
is full.
230 V/AC 50/60 Hz
3500 W/15 A
0.1 -3500 W
0.000 - 9999 kWh
three lines with 4 positions each
0,000 - 9,999
5 - 3500 W (± 1% + 1 count)
2 -5 W (± 5% + 1 count)
< 2 W (±15% + 1 count)
3 V, CR1620
10 - 50 °C/max. 90%rH (not condensing)
Operating altitude: max. 2000 m (above MSL)
ca. 240 g
164 x 82 x 83 (mm)
CAT II
2
Remedy
After the end of the
data tranmissions remove the
SD card from the measurement
device.
Back up the data on a
SD card and delete the
data memory.
37

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis