Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient/ burst
IEC 61000-4-4
Surge IEC
61000-4-5
82
MANUEL UTILISATEUR NEOTENS NOMAD | WWW.SPENGLER.FR
such an environment.
IEC 60601 test
Compliance level
level
± 8 kV
± 8 kV
contact
contact
± 2 kV,±4 kV,
± 2 kV,±4 kV,
± 8 kV,
± 8 kV,
± 15 kV air
± 15 kV air
± 2 kV for
N/A
power supply
lines
± 1 kV for
input/output
lines
± 1 kV
N/A
differential
mode
± 2 kV
common
mode
Electromagnetic
environment
-guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at
least 30 %.