ELECTROMAGNETIC COMPATIBILITY INFORMATION

The device satisfies the EMC requirements of the international standard IEC 60601-1-2. The
requirements are satisfied under the conditions described in the table below. The device is an
electrical medical product and is subject to special precautionary measures with regard to EMC
which must be published in the instructions for use. Portable and mobile HF communications
equipment can affect the device. Use of the unit in conjunction with non-approved accessories
can affect the device negatively and alter the electromagnetic compatibility. The device should
not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Emissions test
Radiated emission
CISPR 11
Conducted emission
CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Compliance
Group 1, class B.
N/A
N/A
N/A
MANUEL UTILISATEUR NEOTENS NOMAD | WWW.SPENGLER.FR
Electromagnetic environ-
ment -guidance
The device uses RF
energy only for its internal
function. Therefore, its
emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment
EN
81