Centrifugation with pre-set time

Adjusting time or recall a programme with pre-set time (see Chapter "Entering centrifugation parameter" or
"Programme recall").
Press the key
. The rotation display
START
After expiration of the time or with truncation of the centrifugation run by pushing the key
effected with the selected brake step. The brake step is displayed.
During the centrifugation run the rotational speed of the rotor or the subsequently resulting RCF value, the sample
temperature (only in centrifuges with cooling) and the remaining time will be displayed.
15.2

Continuous run

Adjusting the symbol ∞ or recall a continuous run programme (see Chapter "Entering centrifugation parameter"
or "Programme recall").
Press the key
. The rotation display
START
at 00:00.
Press the key
in order to stop the centrifugation run. The run-down is effected with the selected brake
STOP
step. The brake step is displayed.
During the centrifugation run the rotational speed of the rotor or the subsequently resulting RCF value, the sample
temperature (only in centrifuges with cooling) and the expired time will be displayed.
15.3

Short-term centrifugation

Hold down the key
IMPULS
Let go of the key
IMPULS
brake step. The brake step is displayed.
During the centrifugation run the rotational speed of the rotor or the subsequently resulting RCF value, the sample
temperature (only in centrifuges with cooling) and the expired time will be displayed.
16

Emergency Stop

Press the key
twice.
STOP
With Emergency Stop the run-down is effected with brake step 9 (shortest run-down time). Brake step 9 is displayed.
If brake step 0 was pre-selected, the run-down time is technically longer than with brake step 9.
17

Acoustic Signal

The acoustic signal sounds:
Upon the appearance of a disturbance in 2 second intervals.
After completion of a centrifugation run and rotor standstill in 30 second intervals.
The acoustic signal is stopped by opening the lid or pressing any key.
The signal after completion of the centrifugation run can be activated or deactivated in the following manner, if the
rotor is at standstill:
Hold down the key
for 8 seconds.
After 8 seconds, SOUND / BELL appears in the display.
Set OFF or ON with the key
Press the key
in order to store the setting.
START
As confirmation, ∗∗∗ ok ∗∗∗ will be displayed for a short period.
18

Recall hours of operation

Recall hours of operation is only possible during rotor standstill.
Hold down the key
for 8 seconds.
After 8 seconds, SOUND / BELL appears in the display.
Press the key
once again.
The centrifuge's hours of operation (CONTROL: ) are displayed.
Press the key
or
to exit the hours of operation recall.
lights up as long as the rotor is turning.
lights up as long as the rotor is turning. The time metering begins
. The rotation display
lights up as long as the rotor is turning.
again in order to stop the centrifugation run. The run-down is effected with the selected
or
.
EN
, the run-down is
STOP
33/80