T/°C
Temperature Set Point (only in centrifuges with cooling). Adjustable from -10°C to +40°C, in 1°C
intervals. The lowest obtainable temperature depends on the rotor (see Chapter "Anhang/Appendix,
Rotoren und Zubehör/Rotors and accessories").
PROG STO Program position on which the program is stored. 4 programs can be stored (program positions 1 - 2 – 3
- 4).
14

Programming

14.1

Programme input/alteration

If no key is pressed for 8 seconds long after the selection or during the input of parameters, the previous
values will be shown in the display. The input of parameter then has to be executed again.
Select the RPM or RCF display with the key
Select the desired parameters using the button
In order to set continuous operation, the parameters t/min and t/sec must be set to zero with the
Continual running is represented in the display by the following symbol, "∞".
The parameter PROG STO can be selected using the button
the knob .
Press the button
START / IMPULS
displayed briefly as confirmation.
The previous data in the program position is overwritten during saving.
14.2

Programme recall

Select the parameter PROG RCL using the button
.
Press the button
START / IMPULS
The parameters can be checked by pressing the button
To leave the parameter display press the button
15

Centrifugation

When the centrifuge is running, according to EN / IEC 61010-2-020, no persons, dangerous substances or
objects may be within the safety margin of 300 mm around the centrifuge.
If the permissible weight difference within the rotor loading has been exceeded, the drive shuts down during
the start-up, the unbalance display lights up, and IMBALANCE is displayed.
A centrifugation run can be stopped at any time by pushing the key
All parameters can be selected and altered during the centrifugation run (see Chapter "Programming").
You can switch-over at any time between the RPM and RCF display with the key
centrifugation radius is necessary if you are working with the RCF display.
If
OPEN
OEFFNEN is displayed, a further operation of the centrifuge is only possible after opening the lid
once.
Turn on the mains switch. Switch position Ι.
Load the rotor and close the centrifuge lid.
15.1

Centrifugation with pre-set time

Adjusting time or recall a programme with pre-set time (see Chapter "Programming").
Press the key
START / IMPULS
After expiration of the time or with truncation of the centrifugation run by pushing the key
down is effected with the selected brake step. The brake step is displayed.
During the centrifugation run the rotational speed of the rotor or the subsequently resulting RCF value, the sample
temperature (only in centrifuges with cooling) and the remaining time will be displayed. After the centrifuge has ended
its run and the rotor has come to a halt, the lid opens automatically on centrifuges without cooling.
. RCF values are displayed in > <.
RCF
and set using the knob
SELECT
in order to store the setting on the desired program position. ∗∗∗ ok ∗∗∗ is
SELECT
. The centrifugation data of the selected program position is displayed.
SELECT
OPEN / STOP
. The LED in the button
START / IMPULS
.
and the desired program position set using
SELECT
and set the desired program position using the knob
.
or press no button for a period of 8 seconds.
.
OPEN / STOP
lights up for as long as the rotor turns.
EN
knob.
. The input of the
RCF
, the run-
OPEN / STOP