4 Getting Started
Please read the Safety Guidelines before using your device, see
Safety Guidelines on page 7 .
This section explains how to start and finish using the device each day. More
detailed instructions can be found in the help files for each application.
4.1 Switching on each day
You must start and close the system each day.
1.
Remove the scan head dust cover.
2.
Press the power switch at the rear of the scan head to the ON ( I )
position.
Note
When switched on from cold, the lasers in the scan head
l
will take a minimum of 15 minutes to warm up to the
operating temperature required for optimum
performance.
You must start the
l
the scan head has been switched on. This is because
both the scan head and the software need to be running
while the system warms up for optimum performance.
An error message may be displayed if capture is
l
attempted before the required temperature has been
reached, see How do I clear a Light Source Error? on page
29 .
3.
Check the Capture PC is switched on. Turn it on if it has been turned
off or has lost power.