All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
►
Follow the instructions for use of individual accessories.
►
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
Coupling the tool to the applied part
Risk of injury when coupling/uncoupling tools in
the "On" position due to accidental activation of
the applied part!
WARNING
►
Only couple/uncouple tools in the "Off" posi-
tion.
►
Lock applied part ("Off" position), see instructions for use of
ELAN 4 electro control unit GA800 (TA014401).
►
Pull back unlocking sleeve 2 on the applied part and push tool shaft 4
in the correct position as far as it will go into the tool coupling 1.
►
Release unlocking sleeve 2.
►
Pull at tool shaft 4 to check that it is securely coupled.
Uncoupling the tool from the applied part
Risk of injury when coupling/uncoupling tools in
the "On" position due to accidental activation of
the applied part!
WARNING
►
Only couple/uncouple tools in the "Off" posi-
tion.
►
Lock applied part ("Off" position), see instructions for use of
ELAN 4 electro control unit GA800 (TA014401).
►
Pull back unlocking sleeve 2 on the applied part and remove tool by the
tool shaft 4 from the tool coupling 1.
6.2
Function checks
►
Prior to each use, check that all products to be used are in good work-
ing order.
►
Check the secure connection of all products to be used.
►
Check that the tool is coupled properly. To do this, pull on the tool.
►
Check that the cutting edges of the tools are not showing any mechan-
ical damage.
►
Briefly run the applied parts at maximum speed.
►
Do not use the product if it is damaged or defective. Set aside the prod-
uct if it is damaged.
6.3
Safe operation
Coagulation of patient tissue or risk of burns for
patients and users from hot applied part/tool!
►
Cool the tool during operation.
WARNING
►
Put down the applied part/tool beyond reach of
the patient.
►
Allow the applied part/tool to cool down.
►
Use a cloth to protect against burns when
changing the tool.
Improper use can result in injury to the dura mater
and brain tissues!
►
DANGER
Apply special care when using perforator driv-
ers.
►
Make certain that the user is familiar with the
skull trepan and its mode of operation.
►
Only use sharp tools.
Risk of infection from aerosol formation!
Risk of injuries caused by particles coming loose
from the tool!
WARNING
►
Take appropriate protective measures, e.g.
watertight protective clothing, face mask and
protective goggles, suction extraction.
Risk of injury and/or malfunction!
►
Always carry out a function check prior to using
WARNING
the product.
It is only possible to operate the applied part and change the setting
parameters on the control unit if:
■
the applied part is connected to the control unit,
■
there is no second applied part released ("On" position) at the same
time and
■
the applied part type 5 is shown in the display of the control unit.
Note
For further information, see instructions for use of ELAN 4 electro control
unit GA800 (TA014401).
7.
Validated reprocessing procedure
7.1
General safety notes
Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for reprocess-
ing.
5