All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
Follow the instructions for use of individual accessories.
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
Coupling the tool to the applied part
Risk of injury when coupling/uncoupling tools in
the "On" position due to accidental activation of
the applied part!
WARNING
Only couple/uncouple tools in the "Off" posi-
tion.
Lock applied part ("Off" position), see instructions for use of
ELAN 4 electro control unit GA800 (TA014401).
Pull back unlocking sleeve 2 on the applied part and push tool shaft 4
in the correct position as far as it will go into the tool coupling 1.
Release unlocking sleeve 2.
Pull at tool shaft 4 to check that it is securely coupled.
Uncoupling the tool from the applied part
Risk of injury when coupling/uncoupling tools in
the "On" position due to accidental activation of
the applied part!
WARNING
Only couple/uncouple tools in the "Off" posi-
tion.
Lock applied part ("Off" position), see instructions for use of
ELAN 4 electro control unit GA800 (TA014401).
Pull back unlocking sleeve 2 on the applied part and remove tool by the
tool shaft 4 from the tool coupling 1.
6.2
Function checks
Prior to each use, check that all products to be used are in good work-
ing order.
Check the secure connection of all products to be used.
Check that the tool is coupled properly. To do this, pull on the tool.
Check that the cutting edges of the tools are not showing any mechan-
ical damage.
Briefly run the applied parts at maximum speed.
Do not use the product if it is damaged or defective. Set aside the prod-
uct if it is damaged.
6.3
Safe operation
Coagulation of patient tissue or risk of burns for
patients and users from hot applied part/tool!
Cool the tool during operation.
WARNING
Put down the applied part/tool beyond reach of
the patient.
Allow the applied part/tool to cool down.
Use a cloth to protect against burns when
changing the tool.
Improper use can result in injury to the dura mater
and brain tissues!
DANGER
Apply special care when using perforator driv-
ers.
Make certain that the user is familiar with the
skull trepan and its mode of operation.
Only use sharp tools.
Risk of infection from aerosol formation!
Risk of injuries caused by particles coming loose
from the tool!
WARNING
Take appropriate protective measures, e.g.
watertight protective clothing, face mask and
protective goggles, suction extraction.
Risk of injury and/or malfunction!
Always carry out a function check prior to using
WARNING
the product.
It is only possible to operate the applied part and change the setting
parameters on the control unit if:
the applied part is connected to the control unit,
there is no second applied part released ("On" position) at the same
time and
the applied part type 5 is shown in the display of the control unit.
Note
For further information, see instructions for use of ELAN 4 electro control
unit GA800 (TA014401).
7.
Validated reprocessing procedure
7.1
General safety notes
Note
Adhere to national statutory regulations, national and international stan-
dards and directives, and local, clinical hygiene instructions for reprocess-
ing.