Herunterladen Diese Seite drucken

Aesculap ELAN 4 Gebrauchsanweisung Seite 6

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 14
Aesculap
ELAN 4 electro perforator driver GA822
4.
Product description
4.1
Scope of supply
Art. no.
Designation
GA822
ELAN 4 electro perforator driver
TA014419
Instructions for use of ELAN 4 electro perforator driver
GA822 (leaflet)
4.2
Components required for operation
Operation-ready ELAN 4 electro control unit GA800, see TA014401
Cranial perforators with HUDSON connection (e.g. GB300R, GB302R,
GB304R)
4.3
Operating principle
The ELAN 4 electro perforator driver can be used to drive cranial perfora-
tor tools.
No wrenches are required for coupling/uncoupling the tools (HUDSON
coupling).
The speed can be adjusted continuously by a manual/foot control.
The recommended speed for cranial perforators with Aesculap tools is
1 200 rpm.
If tools from other manufacturers are used, the manufacturer's instruc-
tions for use must be followed and, if applicable, the tools must be
adjusted to the ELAN 4 electro control unit.
5.
Preparation
Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
Do not use products from open or damaged sterile packaging.
Prior to use, inspect the product and its accessories for any visible
damage.
Use the products and their accessories only if they are in perfect tech-
nical condition.
4
®
6.
Working with the applied part
Risk of infection and contamination!
Applied parts, motor cable and sometimes also
tools are delivered unsterile!
WARNING
Carry out sterile processing of the applied parts,
motor cable and unsterile tools prior to use, in
accordance with the instructions for use.
Risk of injury and material damage due to inadver-
tent activation of the applied part!
WARNING
Secure applied parts that are not actually used
against inadvertent activation ("Off" position),
see instructions for use for ELAN 4 electro con-
trol unit GA800 (TA014401).
Risk of injury and material damage due to inappro-
priate use of tools!
Always follow the safety advice and information
WARNING
given in the instructions for use.
When coupling/uncoupling, handle tools with
cutting edges with care.
Damage to the product if dropped!
Use the products only if they are in perfect
technical condition, see Function check
WARNING
Risk of burns to skin and tissue due to blunt tools/
lack of maintenance of the applied part!
WARNING
Use tools only if they are in perfect condition.
Replace blunt tools.
Ensure correct maintenance of the applied part,
see Maintenance.
6.1
System set-up
Note
Operating elements on the system components of the ELAN 4 electro motor
system are labeled with a gold marking.
Connecting the accessories
The perforator driver is fitted with a HUDSON tool coupling.
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.

Werbung