2. Safety
2.3 Additional safety instructions for instruments per ATEX
EN
The responsibility for classification of zones lies with the plant
■
operator and not the manufacturer/supplier of the equipment.
The plant operator guarantees, and is solely responsible, that all
■
thermometers in use are identifiable with respect to all safety-relevant
characteristics. Damaged thermometers may not be used.
Electrical screening may only be grounded at one end, and outside of
■
the Ex area. Special cases are described in DIN EN 60079-14:2003.
There must be a galvanic separation between the intrinsically safe
■
and the non-intrinsically safe electrical circuits.
2.4 Special hazards
8
WARNING!
Follow the requirements of the ATEX directive.
Additionally the specifications of the respective national
regulations concerning Ex usage (e.g. EN 60079-10 and
EN 60079-14) apply.
Non-observance of these instructions and their contents may
result in the loss of explosion protection.
WARNING!
Neither repairs nor structural modifications are permitted,
and any would void the guarantee and the certification.
WARNING!
Observe the information given in the applicable type
examination certificate and the relevant country-specific
regulations for installation and use in hazardous areas (e.g.
IEC 60079-14, NEC, CEC). Non-observance can result in
serious injury and/or damage to the equipment.
For additional important safety instructions for instruments
with ATEX approval see chapter 2.3 "Additional safety
instructions for instruments per ATEX".
WIKA operating instructions models TR31-3, TR31-K (Ex i)