10. Information on mounting and operation in ...
Thus the entire measuring circuit (including the sensor circuit) is an "ib"
current circuit. Instruments that have been operated in a power supply
circuit of type "ib" cannot be re-used in a power supply circuit of type "ia".
10.2 Overview of the temperature zones
Model TR31-3
■
with transmitter
without neck tube
T
: -40 ... +85 °C
a
Process
connection
T
: -30 ... +150 °C
process
WIKA operating instructions models TR31-3, TR31-K (Ex i)
without transmitter
without neck tube
T
: -50 ... +85 °C
a
Process
connection
T
: -50 ... +150 °C
process
with transmitter
with neck tube
T
: -40 ... +85 °C
a
T
: -40 ... +85 °C
a
Process
connection
T
: -30 ... +250 °C
process
without transmitter
with neck tube
T
: -50 ... +85 °C
a
T
: -50 ... +150 °C
a
Process
connection
T
: -50 ... +250 °C
process
EN
37