RSP3 installation and user's guide
www.renishaw.com
RCP TC, RCP TC-2 and RCP TC-3 LED status
indication and troubleshooting
RCP TC
LED
Green
Amber
Red
RCP TC-2 and RCP TC-3
LED
Green
Flashing green
Amber
Flashing amber
Flashing red
WARNING: The RCP TC, RCP TC-2 or RCP TC-3 plate may be hot.
Issued 10 2019
Status
Power on, probe present
Power on, no probe
Error
Status
Power on, probe present, temperature within
target
Power on, probe present, temperature
outside of target
Power on, probe not present, temperature
within target
Power on, probe not present, temperature
outside of target
Error
Troubleshotting
n/a
If the probe is present, clean the port and
probe contacts.
Cycle power to restart. If the problem
persists contact your supplier.
Troubleshooting
n/a
Wait for the rack to reach stable temperature.
If the probe is present, clean the port and
probe contacts.
If the probe is present, clean the port and
probe contacts. Wait for the rack to reach
stable temperature.
Cycle power to restart. If the problem
persists, contact your supplier.