Herunterladen Diese Seite drucken

VATECH PaX-i Plus PCH-30CS Handbuch Seite 52

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 127
4. Acquiring CEPH Images (Optional)
4.2.1
Lateral / Full Lateral Mode
Patient Positioning
1.
Turn the Nasal Positioner to the Lateral Mode Positioning Marker as shown
below.
2.
Leave enough space between the Ear Rods.
3.
Guide the patient to the CEPH unit.
4.
Direct the patient to relax his/her neck and shoulders and stand upright.
5.
Use the Column UP/DOWN button or switch option to adjust the height of the
CEPH unit to approximately match the height of the patient.
44
Correct posture reduces the shadow casted by the patient's
cervical spine and allows clear image acquisition.
Use the Porion Position Reference Indicator that appears in
the acquired image to easily confirm the location of Porion.
After adjusting the height of the column, align the Ear Rods
and Nasal Positioner to the patient.
PaX-i Plus User Manual