6.
Align the Ear Rods to the patient's ears properly so that the head does not move
during the operation. And align the Nasal Positioner to the patient's nasion by
adjusting its height.
7.
Align horizontally so the patient's Frankfurt Line is parallel with the floor.
8.
Direct the patient to swallow first before closing the mouth and to remain in
his/her current position until image acquisition is completed.
9.
Click the READY button on Console Software. No X-ray will be emitted at this
point.
10. Now go to 4.3 X-ray Exposure to start the exposure.
PaX-i Plus User Manual
4. Acquiring CEPH Images (Optional)
45