compliance level above, Dual Pex should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting of relocating Dual Pex.
** Over the frequency range 150 kHz to 80MHz, fi eld strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the medical device.
Nowadays, many RF wireless equipment's have being used in various healthcare locations where medical equipment and/or systems are used. When they are
used in close proximity to medical equipment and/or systems, the medical equipment and/or systems' basic safety and essential performance may be affected.
The Dual Pex has been tested with the immunity test level in the below table and meet the related requirements of IEC 60601-1-2:2020. The customer and/or
user should help keep a minimum distance between RF wireless communications equipment and the Dual Pex as recommended below.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: at 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: these guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
refl ection from structures, objects and people.
16. Warranty
Inspections
and
repairs
Your product is an accurate medical device, produced under maximum quality and test conditions.
To ensure an individualized follow-up and traceability of your product, please complete the warranty
registration
on-line
on
activate the after-sales procedures for your device so that Micro-Mega can give you full satisfaction.
Micro-Mega guarantees the battery for one year and the device for two years against any hidden manufacturing defect,
excluding faults due to improper use, inappropriate maintenance, accidental damage (fall, impact, etc.), attempted
intervention or repairs by personnel not approved by Micro-Mega, attempted product modifi cation.
17. Statement
Non-contractual pictures. Micro-Mega reserves the right to modify the characteristics of its product without prior notice.
Distributed by Micro-Mega
12, rue du Tunnel
25000 Besancon - France
Caretechion GmbH
Niederrheinstr. 71, 40474
Duesseldorf - Germany
8
Recommended separation distances between portable and mobile RF communications
equipment and the Dual Pex.
150 kHz to 80 MHz
d=1.2√
P
0.13
0.36
1.1
3.5
13
must
only
be
carried
www.micro-mega.com
website
80 MHz to 800 MHz
d=1.2√
P
0.11
0.37
1.3
3.6
10
out
by
an
approved
within
10
days
of
Manufactured by Changzhou Sifary Medical
Technology Co., Ltd.
No. 99, Qingyang Road, Xuejia County, Xinbei
District, Changzhou City, 213000 Jiangsu - China
Year of CE marking: 2017
800 MHz to 2.5 GHz
d=2.3√
P
0.22
0.72
2.1
7.1
20
dealer
or
by
Micro-Mega.
your
purchase.
This
will