Herunterladen Diese Seite drucken

Coltene MicroMega Dual Pex Bedienungsanleitung Seite 6

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 15
Operation
4 Sterilization
5 Storage
• Use only Ethanol for Disinfection (Ethanol 70 to 80% vol.).
14. Main Technical Data
Dimensions - Weight
Display
Power supply
Charger
Power rating
Degree of protection
Electrical safety class - Applied part
Operating conditions
Transport and storage conditions
15. Electromagnetic Compatibility (EMC) – Declaration of conformity
Dual Pex is intended for use in the electromagnetic environment specifi ed below. The customer or the user of Dual Pex shall make sure it is
used in such an environment.
Electromagnetic emission
Emissions test
RF emissions - CISPR 11
RF emissions - CISPR 11
Harmonic emissions IEC61000-3-2
Voltage fl uctuations/fl icker emissions - IEC61000-3-3
6
Concerned parts
Processing
Lip hook (1.C)
Steam sterilization at 135°C at least
File clip (1.B)
4 minutes, or at 121°C at least 35
minutes.
Minimum drying time after sterilization:
10 minutes.
Keep the components in sterilization
packaging in a dry and clean
environment.
20 cm x 17.5 cm x 7 cm (package) - 700g
3.5'' color LCD
Lithium ion battery 3.7V, 1500mAh
AC 100-240V - 50/60Hz - 0.2A
0.3 W
IPX 0
Class II - BF
Indoor use only 10°C / 40 °C - Relative humidity 30%~75% non-condensing at 0° -
Altitude < 2000 m above sea level
-20 °C / +55 °C - Relative humidity 20% - 80%, non-condensing at > 40 °C -
Atmospheric pressure 70 kPa - 106 kPa
Guidance and manufacturer's declaration
Compliance
Group 1
Class B
Class A
Complies
Warning
Use only approved autoclave devices
according to EN 13060 or EN 285.
The sterilization procedure must comply
with ISO 17665.
Waiting for cooling before touching.
Check the packaging before using it
(packaging integrity, no humidity and
validity period), otherwise sterilize again.
Electromagnetic environment - Guidance
Dual Pex uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Dual Pex is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.

Werbung