Electromagnetic immunity
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transients/bursts
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply lines - IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8
Note: UT: rated voltage(s); E.g. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at 60Hz
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF -
3 V
0.15 MHz - 80 MHz,
IEC 61000-4-6
6V in ISM bands
between 0.15 MHz
and 80 MHz, 80% AM
at 1 KHz
Radiated RF -
3 V/m
IEC 61000-4-3
80 MHz to 2.7 GHz,
80% AM at 1 KHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected be absorption and
refl ection from structures, objects and people.
a
Field strengths from fi xed transmitters, such as base stations for ratio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicated theoretically with
accuracy. To assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey
should be considered. If the measured fi eld strength in the location in which Dual Pex is used exceeds the applicable RF
IEC 60601 test level
± 8kV contact
±2, 4, 8k, 15V air
±2 kV
100 kHz repetition frequency
Line to line: ±0.5kV,±1 kV
0% UT; 0.5 cycle at 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0% UT; 1 cycle and 70% UT; 25/30 cycles
sine phase at 0° 0% UT; 250/300 cycle
3 A/m 50Hz or 60 Hz
3V
Portable and mobile RF communications equipment should be used no
0.15 MHz - 80
closer to any part of Dual Pex, including cables, than the recommended
MHz, 6V in ISM
separation distance calculated from the equation applicable to the frequency
bands between
of the transmitter.
0.15 MHz and 80
Recommended separation distance
MHz, 80% AM at 1
d=1.2√
KHz
d=1.2√
d=2.3√
3V/m
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fi xed RF transmitters, as determined
by an electromagnetic site survey, should be less than the compliance level in
each frequency range.**
Interference may occur in the vicinity of equipment marked with the following
symbol:
Compliance level
Electromagnetic environment -
guidance
±8 kV contact
Floors should be in wood, concrete or
±2, 4, 8 k 15V air
ceramic tile. If covered by synthetic
material, the relative humidity should
be at least 30 %.
Mains power quality should be that
±2 kV
of a typical commercial or hospital
100 kHz repetition
frequency
environment.
Mains power quality should be that of a typical
Line to line: ±0.5kV,
commercial or hospital environment.
±1 kV
Mains power quality should be that of a typical
0% UT; 0.5 cycle at 0°, 45°, 90°,
commercial or hospital environment. If the user of
135°, 180°, 225°, 270° and 315°
devices require continued operation during power mains
0% UT; 1 cycle and 70% UT;
interruptions, it is recommended that devices be
25/30 cycles
powered from an uninterruptible power supply or a
sine phase at 0° 0% UT; 250/300
battery
cycle
3 A/m 50Hz or 60 Hz
Power frequency magnetic fi eld
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
P
800 MHz ~ 800 MHz
P
800 MHz ~ 2.5 GHz
P
7