6. Fault / 7. Maintenance, cleaning
Fault
Constant output
signal upon change
in pressure
EN
Mechanical damage
Pairing over
®
Bluetooth
not
working

7. Maintenance, cleaning

Personnel: Skilled personnel
For contact details see chapter 1 "General information" or the back page of
the operating instructions.

7.1 Maintenance

This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
This does not apply to the battery or sensor replacement.
Only use original parts, see chapter 10 "Accessories and spare parts".

7.2 Battery handling

For battery replacement observe the following instructions:
Only change the batteries in a dry environment
Do not use rechargeable batteries
Only use approved batteries, see chapter 7.2 "Battery handling".
Do not use power tools for screwing in.
26
Causes
Mechanical overload caused by
overpressure
Improper handling
Impermissible loading at the
installation location (e.g. fire).
Improper handling (5x incorrect
input of the key)
WIKA operating instructions WIKA-radio unit, model Netris
Measures
Replace instrument; if it fails
repeatedly, contact the manufac-
turer
Replace instrument
Carry out a battery reset and log
in with the correct BLE key
Contact the manufacturer
®