5. Commissioning, operation
Connecting the external power supply
Version with external power supply

5.3 Connecting the radio unit to a measuring instrument

1. Disconnect the radio unit from the external power supply / battery supply after
successful configuration. The selected radio unit configuration is saved.
2. Connect the radio unit to a measuring instrument
→ For the version with an external power supply, connect the measuring instrument to
the end of the connector without colour-coding.
→ For connections with open cable ends, see "Pin assignment of free cable ends" in
chapter 9 "Specifications".
3. Ensure that the power supply to the radio unit is restored. The radio unit restarts with
the settings made previously.
WIKA operating instructions WIKA-radio unit, model Netris
1. Connect the radio unit to the end of the
short Y-cable (marked blue).
2. Connect the (red-marked) connector to
the external power supply.
The supply voltage should not exceed or
drop below the voltage range mentioned
in chapter 9 "Specifications", in order to
ensure trouble-free operation. It may be
necessary to provide a separate fuse for
the radio unit.
The version of the radio unit with an exter-
nal power supply does not have a built-in
battery. Similar to the version with battery,
the case can be opened to check the
status of the LED.
®