Safety precautions

This product complies with the requirements of the following European
Community directives: 2014/30/EU (Electromagnetic Compatibility) and
2014/35/EU (Low Voltage) as amended by 2014/32/EU (CE-Marking).
Overvoltage category III 1000V; pollution degree 2.
CAT I:
For signal level, telecommunication, electronic with small
transient over voltage
CAT II:
For local level, appliances, main wall outlets, portable equipment
CAT III:
Supplied from a cable under earth; fixed installed switches,
automatic cut-off or main plugs.
CAT IV:
Units and installations, which are supplied over aerial line, which
are stand in a risk of persuade of a lightning, i.e. main-switches
on current input, overvoltage-diverter, current use counter.
To ensure safe operation of the equipment and eliminate the danger of
serious injury due to short-circuits (arcing), the following safety precautions
must be observed.
*
Do not use this instrument for high-energy industrial installation
measurement.
*
Do not exceed the maximum permissible input ratings (danger of
serious injury and/or destruction of the equipment).
*
The meter is designed to withstand the stated max voltages. If it is not
possible to exclude without that impulses, transients, disturbance or for
other reasons, these voltages are exceeded a suitable prescale (10:1)
must be used.
*
Disconnect test leads or probe from the measuring circuit before
switching modes or functions.
*
Never touch the tips of the test leads or probe.
*
Do not conduct voltage measurements with the chosen measurement
range of µA or mA.
-17-