Herunterladen Inhalt Inhalt Diese Seite drucken

Product Selection; Fitting Instructions - Ossur MIAMI J Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für MIAMI J:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
Warning: Patient should immediately contact a healthcare professional:
– If there is a change or loss in device functionality, or if the device
shows signs of damage or wear hindering its normal functions;
– If any pain, skin irritation, or unusual reaction occurs with the use
of the device.
Warning: The device should be used directly on intact skin and never an
open wound. Inspect the contact area before and after each use.
The device is for single patient – multiple use.

PRODUCT SELECTION

Miami J Collar Sizes (Fig. 2)
Miami J collar sizes are phenotype-driven. To determine the appropriate
collar size, identify your patient´s phenotype according to the Sizing
Silhouettes (Fig. 2) & Simple Sizing Questions below.
Specialty sizes < 40% of Population
a. 200S - Super Short/Kyphotic: Does your patient have a kyphotic(chin-
on-chest) neck?
Kyphotic / Chin-on-chest / Geriatric / Osteoporitic and/or ankylosing
spondylitic.
b. 200L - Stout: Does your patient have a very large neck circumference?
Very large circumference / Obese / "No-neck" / Massive shoulders.
c. 250 - Xs: Does your patient have a very short, thin neck?
Very short thin neck / Principally female / More prevalent in the Asian
population / Rare.
d. 500 - Tall: Does your patient have a long tall neck?
Long, tall neck / "Swan neck" / Young women / Adolescents.
Standard sizes > 60% of Population
e. 400 - Regular: Is your patient female?
Female / Standard size for mature women / Thin, mature men.
f. 300 - Short: All other adults.
Standard size for adult men / Short-necked women.
Quick Size Check (Optional) Fig. 3
Use your fingers to measure the vertical chin-to trapezius distance.
On the collar, this measurement corresponds to the distance from the
bottom of the chin support to the bottom edge of the collar plastic.
NOTE: Proper sizing is critical to achieving comfort and proper
immobilization.

FITTING INSTRUCTIONS

Supine Application
Ensure that correct spinal protocol is followed throughout the application.
1. Slide the Back behind the patient´s neck and center it, noting arrow
on panel pointing up (Fig. 4). Ensure the Sorbatex™ padding extends
beyond the edge of the plastic.
NOTE: Long hair should be placed outside of the plastic.
2. Flare the sides of the Front out, slide it up the chest wall and scoop it
up under the chin (Fig. 5). Sides of the collar Front should be oriented
up, off the trapezius, and toward the ears.
3. While holding the Front securely, curl the ends snugly against the
patient´s neck (Fig. 6). With the Front of the collar inside the Back,
apply the Hook and Loop Strap and secure the opposite side of the
collar in the same fashion. Tighten straps alternately to an equal
length on both sides.
6