26 Close-quartered Room Conditions
26 Close-quartered Room Conditions
If the prevailing room conditions do not permit full
movement of the assistant element support arm, limit its
travel range as follows:
1.
Set the support arm to the position next to the cuspidor.
2.
Unscrew the two screws of the locking device.
3.
Rotate the locking device 180° and fasten it again with
the two screws.
•
Now the support arm can only be swiveled. The height
of the assistant element is no longer adjustable.
34
2.
3.
Sirona Dental Systems GmbH
1.
59 06 602 D 3382
D 3382.031.01.02.02