20 Comparing and Setting Operating Air Pressures

20 Comparing and Setting Operating Air Pressures
TL
i
NOTE
The operating pressures for the instruments are set at the
factory. The settings for the driving air and spray air can be
changed at the customer's request.
The driving air and spray air pressures can be set exactly
using the four-way measuring head (Order No. 36 66 336).
Factory settings with turbine running:
Turbines:
T1 Control, TC1 and TC3, TM1 TITANIUM:
Driving air on four-way measuring head (TL)
2.7 bar ± 0.1 bar
Spray air on four-way measuring head (SL)
2.7 bar ± 0.1 bar
26
SW
EL 1, motor SL
The cooling air pressure for the motor results from the
driving air setting of the turbine. A turbine must be fitted
instead of the motor to perform the setting (see page 27 for
setting).
i
NOTE
TL = driving air
SW = spray water
SL = spray air
RL = return air
Sirona Dental Systems GmbH
SL
RL
59 06 602 D 3382
D 3382.031.01.02.02