Connecting the Air, Water and Suction Hoses

Air
1.
Water
Air
1.
ATTENTION
Do not damage sanitary hoses and pipes.
ATTENTION
Make sure they are connected properly!
ATTENTION
The sanitary hoses must not contact belt band A!
1. Connect flexible connecting pipes for water (green
marking) and air (blue marking) to the on-site corner
valves.
Screw seals tight with nuts.
2. Plug in the end piece of the drain hose (smaller hose
diameter).
ATTENTION
Check that the suction pipe is connected as a suction pipe
and is not confused with the drain!
59 06 602 D 3382
D 3382.031.01.02.02
A
3.
2.
ATTENTION
Make sure the drain slopes over its whole length. Do not run
the hose over other hoses, prevent siphoning.
is not required if the water is removed centrally by suction. If
the drain pipe is present, you must cover it.
3. Plug in the end piece of the suction hose.
ATTENTION
Use only suitable wrenches for the installation. Never use
pipe wrenches.
Observe the permissible bending radius and avoid sharp
bends.
Never twist the hoses. Also make sure that unintentional
twisting does not occur when tightening down. Use an
additional wrench to prevent any turning.
Tensile forces must not act on the connecting hoses. Make
sure that tensile stressing does not occur after installation
as well.
8 Connecting the Air, Water and Suction Hoses
right
wrong
right
wron
g
The drain
13