Fastening the Chair Base to the Floor

2.
Ø 12mm
X
7mm
Y
160
mm
3.
80
mm
min.
Z
20mm
1. Line the installation template up to the corner valves
and align it dimensionally.
Center punch fastening points A to G.
For left-handed version: H and H' instead of C and C'.
ATTENTION
Fastening points D and E bear the main load of the floor
fastening!
ATTENTION
Drilled hole depth: The screed X and footfall sound
absorption layer Y must be overcome and the concrete Z
must be drilled down to a depth of at least 80 mm for secure
anchoring.
ATTENTION
Floor heating! Determine the position of the heating coils
with the heating turned on.
59 06 602 D 3382
D 3382.031.01.02.02
M
C'
C
A
H'
H
B
ATTENTION
Do not use plugs for wooden floors. With wooden floors
provide appropriate supporting beams and make the
necessary arrangements. With wooden floors drill to a
diameter of ∅7 mm and a depth of 80 mm.
2. Drill the plug holes.
3. Insert the plugs into the supporting concrete.
4. Grasp the chair in area M, lift it off the pallet and move it
into position.
ATTENTION
Do not lift the chair by its backrest or by the bracket for the
upholstery.
7 Fastening the Chair Base to the Floor
M
F
G
1.
D
E
4.