10.7 BATTERY CARE AND MAINTENANCE

Lithium-ion batteries require special care to ensure proper performance and long life. Use only compatible batter-
ies with your device.
• Keep Dry: Always keep liquids away from batteries. If batteries become wet, discontinue use immediately and
dispose of battery properly.
• Effect of temperature on battery performance: The battery powers the device under most environmental con-
ditions. To extend the run-time of your battery, avoid running in temperatures less than 41˚F (5˚C) or higher than
95˚F (35˚C) for extended periods of time.
• Battery Storage: Remove your battery from the device when it is not in use to avoid inadvertent discharge.
Store battery in a cool, dry place. Store with a charge of at least 40-50%. Batteries should be charged up to full
charge and discharged down to 0% at least once every 90 days to maintain maximum lifetime. Avoid storing
your device Battery in extreme temperatures, below -4˚F (-20˚C) or above 140˚F (60˚C), for any amount of time.
• Battery Disposal: Contact your provider for proper battery disposal. Lithium-ion batteries, like all rechargeable
batteries, are recyclable and should never be incinerated.

10.8 SERVICE LIFE

The expected service life of the device is 5 years, except for the sieve beds (plastic columns) which have an
expected life of 1 year and the batteries, which have an expected life of 500 full charge/discharge cycles.

11. PAIRING YOUR DEVICE WITH THE CONNECT APP

The Inogen Connect App pairs your portable oxygen concentrator to your mobile device or tablet using Bluetooth
technology. It is not available in every country – contact your equipment provider for more information.
IMPORTANT: the app is not intended to replace the user interface panel, which is the primary source of information
to which the patient should refer when operating the device.
IMPORTANT: Connection of the Inogen Rove 6 to a Bluetooth connection that includes other equipment could result
in previously unidentified risks to patients, operators or other third parties. The responsible organization should
identify, analyze, evaluate and control these risks. Subsequent changes to the Bluetooth connection could intro-
duce new risks and require additional analysis. Changes to the Bluetooth connection include:
• Changes in the Bluetooth configuration.
• Connection of additional items to the Bluetooth connection.
• Disconnecting items from the Bluetooth connection.
• Update of equipment connected to the Bluetooth connection.
• Upgrade of equipment connected to the Bluetooth connection.