The following figure shows the overall flange di-
mensions with the protection screen fitted on
pump with ISO 200F flange and pump with ISO
200K flange.

WATER COOLING KIT INSTALLATION

Two types of water cooling kits are available to be
mounted when the pump is used under heavy load
conditions or when air cooling is insufficient.
The two model part numbers are: 969-9337 (me-
tallic model), and 969-9347 (plastic model).
Model 969-9337
Model 969-9347
The items of the plastic model kit must be assem-
bled as shown in the following figure
The assembled kit must be screwed into the suit-
able holes of the pump body with a recommended
closing torque of 5 Nm.
Two 6 mm (1/4") internal diameter rubber or plastic
hoses from the water supply must be fitted to the
two nozzles.
These hoses must be held on the respective noz-
zles using hose clips to avoid that the tube(s) gets
loose or disconnected during operation.
Cooling may be carried out either through an open
circuit with eventual discharge of the water, or us-
ing a closed circuit cooling system.
81/87