4. Acquiring CEPH Images (Optional)
4.2.4
Waters' view Mode
Patient Positioning
1.
Turn the Nasal Positioner to the PA / Waters' view / Carpus mode positioning marker
as shown below.
2.
Fold the Nasal Positioner up. The Nasal Positioner is not used in Waters' view mode.
3.
Guide the patient to the CEPH unit.
4.
Ask the patient to stand upright facing the sensor. Make sure that the patient's
shoulders are level and that his/her neck is relaxed.
5.
Use the Column UP/DOWN button or switch option to adjust the height of the CEPH
unit to approximately match the height of the patient.
6.
During the operation, properly align the Ear Rods to the patient's ears so his/her head
does not move.
7.
Direct the patient to swallow first before closing his/her mouth, and guide the patient to
bend the head backward 30˚ - 40˚. Direct the patient to remain in the current position
until image acquisition is completed.
8.
Click the READY button. The x-ray exposure has not started yet.
9.
Now go to 4.3 X-ray Exposure to start the exposure.
60
After adjusting the height of the column, align the Ear Rods to the
patient.