Sterilising
CAUTION
Risk of infection by residual germs
If there is dirt on the parts, germs capable of reproduction
may remain despite the sterilisation process. As a result,
there is a danger of infection.
• Clean, disinfect, and dry all parts thoroughly before steril-
ising.
• Use only validated procedures for cleaning and disinfec-
tion.
EQUIPMENT:
The method has been validated in Europe using:
– Steam steriliser with fractionated pre-vacuum in accordance
with DIN EN 285 or DIN EN 13060
– Sterile barrier system in accordance with DIN EN 11607-1
– Mask stabiliser for the PARI adult mask soft
– Temperature: 132 °C/134 °C
– Holding time: min. 3 minutes
4) Mask stabilisers can be ordered from PARI GmbH separately.