PROCEDURE:
If a disinfectant without a cleaning function is used, the follow-
ing procedure must be performed twice: the first time with a
cleaning agent, and then with the disinfectant.
To ensure safety when handling chemicals, follow the instruc-
tions for use of the agent used, particularly the accompanying
safety instructions.
• Prepare the solution in the basin in accordance with the
manufacturer's instructions. The quantity of solution required
depends on the number of parts that are to be cleaned.
• Place all the individual parts in the prepared solution and
leave them to soak.
Info: If the application period is exceeded significantly, the
plastic parts may take on the smell of the disinfectant.
RINSING:
CAUTION
Risk of allergic reactions and irritation of the mucous
membrane by disinfectants
Disinfectants can trigger allergic reactions or irritation of the
mucous membrane on contact with the skin.
• Rinse the product thoroughly to ensure that no residues of
the disinfectant remain on the PARI product.
• Rinse off all parts thoroughly in running drinking water.
Water temperature: approx. 15 °C, duration: 3 minutes
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean, and absorbent surface, and
allow them to dry completely.