Herunterladen Inhalt Inhalt Diese Seite drucken

Pari VORTEX 051 Gebrauchsanweisung Seite 88

Vorschau ausblenden Andere Handbücher für VORTEX 051:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Chemical
cleaning with
disinfection:
EQUIPMENT:
The method has been validated in Europe using:
– Aldehyde-free instrument disinfectant: Bode
®
Bomix
 plus (concentration: 2%)
Active agent basis: Quaternary ammonium
compound
– Drinking water at about 15 °C
– Application time: 5 minutes
PROCEDURE:
• Clean and disinfect the individual parts in a
single work step, using a solution prepared
according to the manufacturer's instructions.
Info: If the recommended treatment time is
exceeded significantly, the plastic parts may
take on the smell of the disinfectant.
RINSING:
 CAUTION! Disinfectant residues can
cause allergic reactions or irritation of the mu-
cous membrane. Rinse off all parts thoroughly
in running drinking water at about 15 °C for 3
minutes.
DRYING:
• Shake the water out of all of the parts.
• Place all parts on a dry, clean, and absorbent
surface, and allow them to dry completely.
– 40 –

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis