10. To continue recording, hold down the REL/PC (Log/◄) button again for approx. 2 s.
11. To exit the SHIFT mode, press the SHIFT/SETUP button.
Settings for the data memory:
1. Make the settings for the data memory in the SETUP menu.
2. With the measuring device switched on, hold down the SHIFT/SETUP button (3) for approx. 2 s.
will appear on the display. You are now in SHIFT mode.
3. Hold down the SHIFT/SETUP button again for approx. 2 sec to open the SETUP menu.
4. Now press the SHIFT/SETUP button several times until the menu item "LOGGER MEMORY" is
selected. An asterisk symbol * to the left of the menu item shows that the menu item is selected.
5. Press the REL/PC (Log/◄) or MAX/MIN (View/►) button to select "FIX" or "OVERWRITE".
6. The setting "FIX" means that the measured values will be stored until the memory is full (max. 1000
measured values). The measurement will be stopped at the memory number "1000".
7. The setting "OVERWRITE" means that the measured values will be continually stored. If the
measured value "1000" is reached, the device will begin to overwrite the measured values. The
memory number will no longer be displayed. Instead, a blinking infinity symbol "∞" will appear.
8. The menu item "LOGGER MEMORY" is preset to "FIX" by default.
9. Hold down the SHIFT/SETUP button for approx. 2 sec to save the setting and exit the SETUP
menu. You will also exit SHIFT mode.
Display setting for automatic measured value recording:
1. In the automatic measurement, it is possible to activate the energy saving mode.
2. Make these settings in the SETUP menu.
3. With the measuring device switched on, hold down the SHIFT/SETUP button (3) for approx. 2 s.
will appear on the display. You are now in SHIFT mode.
4. Hold down the SHIFT/SETUP button again for approx. 2 sec to open the SETUP menu.
5. Now press the SHIFT/SETUP button several times until the menu item "LOGGER DATA DISPLAY"
is selected. An asterisk symbol * to the left of the menu item shows that the menu item is selected.
6. Press the REL/PC (Log/◄) or MAX/MIN (View/►) button to select "ON" or "OFF".
7. The "ON" setting means that the display will always be switched on during the automatic
continuous measurement.
71