Herunterladen Inhalt Inhalt Diese Seite drucken

Benning MM 11 Bedienungsanleitung Seite 29

Digital-multimeter
Vorschau ausblenden Andere Handbücher für MM 11:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 7
Measuring range Frequency range
20 mV, 200 mV
2 V, 20 V
200 V
750 V
Maximum resolution: 1 µV within the 20 mV measuring range
*
In the frequency range 5 kHz ~ 50 kHz 20 digits must be added to the speci-
1
fied measuring range limit value - for display values less than 50 % of the
measuring range limit value.
Within a frequency range of 50 kHz ~ 100 kHz, display values less than
40 % of the measuring range limit value are not specified.
7.3 Direct current ranges (Switch setting: A DC, mA DC)
overload protection:
-
F 1 A (600 V) fuse, 10 kA, (Bussmann BBS-1 or equivalent) on the mA
input,
-
F 15 A (600 V) fuse, 100 kA, (Bussmann KLK-15 or equivalent) 10 A on the
input,
Measuring range Resolution
20 mA
200 mA
2 A
10 A
7.4 Alternating current ranges (Switch setting: A AC, mA AC)
The measured value is obtained and displayed as a real effective value (TRUE
RMS).
Selectable coupling type: AC or AC+DC. For the coupling AC+DC an additional
error of 1 % + 80 Digit must be taken into account. In the case of non-sinus-
shaped curves the display value becomes less precise. Thus an additional error
results for the following crest factors:
Crest factor from 1.4 to 3.0 additional errors + 1.5 %
Crest factor from 3.0 to 4.0 additional errors + 3.0 %
Overload protection:
-
F 1 A (600 V)-fuse, 10 kA, (Bussmann BBS-1 or equivalent) on the mA
input,
-
F 15 A (600 V)-fuse, 100 kA, (Bussmann KLK-15 or equivalent) 10 A on the
input,
Measuring range Frequency range
20 mA
200 mA
2 A
10 A
03/ 2011
40 Hz ~ 100 Hz
± (0.7 % of the measured value + 80 Digit)
100 Hz ~ 1 kHz
± (1.0 % of the measured value + 80 Digit)
40 Hz ~ 100 Hz
± (0.7 % of the measured value + 50 Digit)
100 Hz ~ 1 kHz
± (1.0 % of the measured value + 50 Digit)
1 kHz ~ 10 kHz
± (2.0 % of the measured value + 60 Digit)
10 kHz ~ 20 kHz
± (3.0 % of the measured value + 70 Digit)
20 kHz ~ 50 kHz
± (5.0 % of the measured value + 80 Digit)
50 kHz ~ 100 kHz ± (10.0 % of the measured value + 100 Digit)
40 Hz ~ 100 Hz
± (0.7 % of the measured value + 50 Digit)
100 Hz ~ 1 kHz
± (1.0 % of the measured value + 50 Digit)
1 kHz ~ 10 kHz
± (2.0 % of the measured value + 60 Digit)
10 kHz ~ 20 kHz
± (3.0 % of the measured value + 70 Digit)
20 kHz ~ 50 kHz
± (5.0 % of the measured value + 80 Digit)
40 Hz ~ 100 Hz
± (0.7 % of the measured value + 50 Digit)
100 Hz ~ 1 kHz
± (1.0 % of the measured value + 50 Digit)
1 µA
± (0.2 % of the measured value + 40 Digit)
10 µA
± (0.2 % of the measured value + 40 Digit)
100 µA
± (0.2 % of the measured value + 40 Digit)
1 mA
± (0.2 % of the measured value + 40 Digit)
40 Hz ~ 500 Hz ± (0.8 % of the measuring range + 50 Digit)
500 Hz ~ 1 kHz
± (1.2 % of the measured value + 80 Digit)
40 Hz ~ 500 Hz ± (0.8 % of the measuring range + 50 Digit)
500 Hz ~ 1 kHz
± (1.2 % of the measured value + 80 Digit)
1 kHz ~ 3 kHz
± (2.0 % of the measured value + 80 Digit)
40 Hz ~ 500 Hz
± (0.8 % of the measured value + 50 Digit)
500 Hz ~ 1 kHz
± (1.2 % of the measured value + 80 Digit)
40 Hz ~ 500 Hz
± (0.8 % of the measured value + 50 Digit)
BENNING MM 11
Meas. precision
Meas. precision
Meas. precision
Overload protection
*
1
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
750 V
Voltage drop
800 mV max.
800 mV max.
1 V max.
1 V max.
Voltage drop
800 mV max.
800 mV max.
1 V max.
1 V max.
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
eff
23

Quicklinks ausblenden: