Herunterladen Inhalt Inhalt Diese Seite drucken

Electrical Connections - WIKA CPH6200-S1 Betriebsanleitung

Hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH6200-S1:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
For gauge sensors, the pressure compensation vent hole is found
at the top of the sensor housing. This vent (with integral membrane
seal) must be kept clear without fail.
Maintenance advice
Both the CPH6200 and sensors are manufactured using solidstate
technology, and contain no moving parts which could wear. If the
instrument housing has been opened, the warranty becomes invalid.
If cleaning is necessary, use a cloth wetted with mild detergent. Avoid
any solvents, caustic or abrasive substances.
As is normal for all measuring instruments, the accuracy of the device
should be tested at regular intervals (approx. once per year - see
Section 7).

1.3 Electrical connections

On the upper edge of the device:
The connector sockets for attaching the CPH series pressure sensors,
CH1 and CH2 (CH2 only with the 2-channel version), are located on the
top edge of the CPH6200 (see Section 5), as is the socket for the WIKA
interface cable (see Section 4).
Operation as analogue output: connection via suitable cable.
The output mode has to be configured via menu and influences
battery life!
On the left side of the device:
The mains adapter socket for connecting the power supply unit (see
current CPH6200 price list) is located on the left side of the handheld
unit.
WIKA Operating Instructions Hand-Held Pressure Indicator
CH1 CH2 (only for 2-channel version: CPH6200-S2)
Connection for WIKA interface cable
or optional analogue output
GB
7

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cph6200-s2Cph6k2guqCph6200qCph6200

Inhaltsverzeichnis