When using the Overlay Display mode, the instrument adjusts the horizontal and trigger settings of the trigger source channel to
display a triggered waveform for that channel. Vertical scale and position adjustments for all active channels in Overlay Display mode
are controlled by the Autoset in Overlay Display Mode Optimizes selection in the Autoset panel of the User Preferences menu. If
the selection is Visibility, Autoset vertically scales and positions all active channel waveforms such that they are uniformly spaced on
screen. If the selection is Resolution, Autoset vertically scales and positions all active channel waveforms such that they each use as
much of the ADC's range as possible.
Note: You can set which parameters the instrument can adjust when running an Autoset. Access the Autoset panel in Utility >
User Preferences > Autoset. See
Autoset guidelines
Autoset displays three or four cycles (depending on the detected signal) with the trigger level near the midlevel of the signal.
The trigger is set to type Edge, rising slope, DC coupling.
If no channels are displayed before pushing Autoset, the oscilloscope adds Ch 1 to the Waveform view whether it has a signal or not.
Autoset ignores math, reference, and bus waveforms.
A channel or waveform with a frequency less than 40 Hz is classified as no signal.

How to trigger on a signal

Use this procedure to open the Trigger menu to select and configure the trigger event type and conditions.
1. Double-tap the Trigger badge on the Settings bar to open the Trigger configuration menu.
2. Select a trigger from the Trigger Type list. The trigger type sets what fields are available in the menu and also updates the illustration
to show a graphic of the trigger type.
To trigger on a bus, you must first add the bus to the Waveform view. See
Note: Triggering on buses other than Parallel requires purchasing and installing serial trigger and analysis options.
3. Select the other fields and panels to refine the trigger conditions. The menu fields and trigger graphic update as you make changes to
the trigger settings. Displayed fields depend on the selected trigger type. Selection changes take effect immediately.
5 Series B Mixed Signal Oscilloscopes MSO54B, MSO56B, MSO58B Quick Start Manual
User Preferences (Utility
menu).
Add a math, reference, or bus waveform
Operating basics
on page 49
47