Use this procedure to compensate supported TPP-family probes that show a Default status when connected to the oscilloscope.
Note: Doing a Default Setup does not delete probe compensation values. A factory calibration does delete all stored probe
compensation values.
Prerequisite: The oscilloscope must be powered on for at least 20 minutes before compensating a probe.
1.
Connect a supported probe to an input channel.
2.
Connect the probe tip and ground lead of the probe to the PROBE COMP terminals on the lower right of the oscilloscope (see
following image).
Connect the probe tip to the 1 kHz source, and the ground clip to the ground. For best results, remove any probe tip accessories and
hold the probe tip directly onto the 1 kHz connector.
Note: Connect only one probe at a time to the PROBE COMP terminals.
3.
Turn off all channels.
4.
Turn on the channel to which the probe is connected.
5.
Push the front-panel Autoset button. The screen displays a square wave.
6.
Double-tap the badge of the channel that you want to compensate.
7.
Tap the Probe Setup panel.
If the Probe Compensation Status says Pass, the probe is already compensated for this channel. You can move the probe to another
channel and start again from step 1, or connect a different probe to this channel and start from step 1.
If the Probe Compensation Status says Default, continue with this procedure.
8.
Tap Compensate Probe to open the Probe Compensation dialog.
9.
Tap Compensate Probe to run the probe compensation.
10. The probe compensation is finished when the Probe Compensation Status displays Pass. Disconnect the probe tip and ground from
the PROBE COMP terminals.
11. Repeat these steps for each supported passive probe that you want to compensate for this channel.
12. Repeat these steps to compensate supported probes on other channels of the oscilloscope.
Note: For most accurate measurements, open the Probe Setup panel and verify the Probe Compensation Status is Pass
whenever you attach a probe to a channel.

Connect to a network (LAN)

Connecting to a network allows you to remotely access the instrument.
Work with your network administrator to obtain the required information to connect to your network (IP address, Gateway IP address,
Subnet Mask, DNS IP address, and so on).
5 Series B Mixed Signal Oscilloscopes MSO54B, MSO56B, MSO58B Quick Start Manual
Configure the instrument
42

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Mso54bMso56bMso58b

Inhaltsverzeichnis