This instrument shows a warning triangle symbol and the words Clipping in a Channel badge when a vertical clipping condition exists. Any
measurement badges associated with that channel also indicate a clipping condition by turning the measurement text red and listing the
type of clipping (positive or negative).
To close the clipping message, change the vertical scale to show the entire waveform, disconnect the probe tip from the excessive voltage
source, and check that you are probing the correct signal using the correct probe.
Clipping causes inaccurate amplitude-related measurement results. Clipping also causes inaccurate amplitude values in saved waveform
files. If a math waveform is clipped, it will not affect amplitude measurements on that math waveform.
Error Messages and Badges
This instrument shows a warning triangle symbol and an error message abbreviation in a Channel badge when an error occurs.
To remove the message from the badge, clear the error as indicated in the table.
Table 3: Probe errors
Error message
Prb Comm
Prb ROM
Unsup
Prb Fault
Over Rng
Temp
No Tip
Tip Fault
S-param
History badge
The History badge is shown in the Results bar. Navigate the history acquisitions by using the Previous/Next or Play/Pause buttons.
Timestamp shows the time difference between the previous acquisition and the selected acquisition.
5 Series B Mixed Signal Oscilloscopes MSO54B, MSO56B, MSO58B Quick Start Manual
Description
Accessory communication timed out. Please re-attach the accessory.
Unable to read probe ROM. Please re-attach the accessory.
Accessory is unsupported.
Critical accessory fault. Please re-attach the accessory. If the problem persists, contact
Tektronix service.
The signal voltage or current is over range. Please reduce the signal amplitude.
The probe has experienced an over temperature condition. Please remove the probe from
the high temperature area.
No probe tip detected. Please install a compatible probe tip.
The probe tip has a fault. Please remove and replace the probe tip.
Error during S-parameter transfer. Please reattach the probe. If the problem persists,
contact Tektronix Service.