Operation

Before first use

READ INSTRUCTION MANUAL BEFORE FIRST USE!
The first cycle must be run without load; with water only.

Loading the autoclave

• Fill the chamber with 1.5 liters of distilled water.
• The device can also be filled with a larger amount of water. However, this leads to long heating times
and very slow cooling.
• In order to avoid calcification of your device, we recommend that you use distilled water and decalcify
the device regularly.
• Place the instrument plate in the chamber and put the load in the chamber. It is important that the
steam can circulate freely, therefore the load should be at a distance from the autoclave chamber.
Attention: Do not autoclave tightly sealed containers (especially glass bottles), since there is no pressure
compensation.

Close and Locking

Check if the gasket is inserted in the lid.
Press the red knob and turn it clockwise to OPEN.