Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
The I-TECH UE is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such an
environment
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
overvoltage
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supplì input
lines
IEC 61000-4-11
IACER Srl
ALL EQUIPMENT AND SYSTEMS
Test level
IEC 60601
±8kV in contact
±15kV on air
±2kV
for power
supplies lines
±1kV
for input/output
lines
±1kV line(s) to
line
±2kV line(s) to
earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycles
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5s
56
Compliance level
Floors should be
wood, concrete or
ceramic tile. If floors
±8kV in contact
are covered with
±15kV on air
synthetic material,
the
humidity should be
at least 30%.
±2kV
Mains power quality
for power
should be that of a
supplies lines
typical commercial
±1kV
or
for input/output
environment.
lines
±1kV line(s) to
Mains power quality
line
should be that of a
typical commercial
±2kV line(s) to
or
earth
environment.
<5% U
Main power quality
T
(>95% dip in U
)
should be that of a
T
for 0,5 cycles
typical commercial
or
40% U
environment. If the
T
(60% dip in U
)
user of I-TECH UE
T
for 5 cycles
requires continued
operation
70% U
power
T
(30% dip in U
)
interruptions, it is
T
for 25 cycles
recommended that
I-TECH
<5% U
powered from an
T
(>95% dip in U
)
uninterruptible
T
for 5s
power supply (UPS)
or a battery.
Electromagnetic
environment -
guide
relative
hospital
hospital
hospital
during
mains
UE
be
MNPG183-02