• If the LifePad
®
the defibrillator.
• Please report any unexpected operation or events to the manufacturer.
• Device not suitable for use in presence of anaesthetic mixtures inflammable with air, oxy-
gen or nitrogen protoxide.

General precautions

• The LifePad
is intended for single use only and is therefore not suitable for reuse:
®
- After a single use and the associated closing of the circuit through pulling the battery
strip, the supplied battery is discharged, meaning that it can no longer be ensured that
the product can be used again at a later time.
- Furthermore, the product cannot be used on different people for hygienic reasons, as it
cannot be disinfected or cleaned.
• Protect the LifePad
tuations, direct sunlight, water, sand and sources of heat that are too close to it (ovens or
heaters), as faultless functionality of the LifePad
• Under no circumstances should you open or repair the LifePad
ality can no longer be guaranteed thereafter and you may receive an electric shock. The
LifePad
cannot be maintained and/or calibrated.
®
• Do not use additional parts that are not recommended by the manufacturer or offered as
accessories.
• If you see or suspect damage to the LifePad
• The LifePad
is protected against splashes. Never submerge the LifePad
®
Notes on handling batteries
• If your skin or eyes come into contact with battery fluid, wash the affected area with water
and seek medical assistance.
Choking hazard! Small children may swallow and choke on batteries. Therefore, batter-
ies should be stored out of the reach of small children.
• Observe the plus (+) and minus (-) polarity signs.
• If a battery has leaked, put on protective gloves and clean the battery compartment with
a dry cloth.
• Protect batteries from excessive heat.
Risk of explosion! Do not throw batteries into a fire.
• Do not charge or short-circuit the batteries.
• Do not use rechargeable batteries.
• Do not disassemble, open or crush the batteries.
• Battery must comply with IEC 60086-4.

5. STORAGE

Possible damage due to incorrect storage
• Store the LifePad
• Make sure to store the product in a place that is quick and easy to access in case of an
emergency.
• The LifePad
must not be bent or subjected to other external stresses during storage, as
®
this could damage the pressure sensor.
is used in combination with a defibrillator follow the safety instructions of
against impacts, moisture, dust, chemicals, strong temperature fluc-
®
in a dry environment with no external stresses.
®
can no longer be guaranteed thereafter.
®
, you must replace it with a new LifePad
®